 AC-DC Medical 25 Watts PBM25 Series
*
THE XPERTS IN POWER
Small Size 2.50" x 4.25" x 0.96" * Low Leakage Current * Medical Safety Agency Approvals * Both Line and Neutral Fused * Meets Class B, EN55011 Conducted * Fixed Switching Frequency
Specification
Input
Input Voltage Input Frequency Input Current Inrush Current Earth Leakage Current * * * * * 85-264 VAC 47-63 Hz 0.7 A (rms) at 115 VAC, 0.4 A (rms) at 230 VAC 10 A max at 115 VAC, 25 A max at 230 VAC 90 A max at 115 VAC 60 Hz 150 A max at 230 VAC 50 Hz
General
Efficiency Isolation * * 68% minimum for 5 V & multi output models, 70% minimum all others 4000 VAC Input to Output 1500 VAC Input to Ground 500 VAC Output to Ground 42 KHz 5 KHz 500,000 hrs per min MIL-HDBK-217F 4.25 x 2.50 x 0.96" 107.95 x 63.50 x 24.38 mm Single output models 163 g, Multi output models 175 g
Switching Frequency * MTBF * Size * Weight *
Output
Output Voltage Output Power Minimum Load Start Up Delay Start Up Rise TIme Hold up Time Initial Set Accuracy Line Regulation Total Regulation Transient Response Ripple & Noise Overvoltage Protection Overcurrent Protection Short Circuit Protection Temperature Coefficient * * * * * * * * * * * * 5.1 to 28 VDC, See Table 25/28 Watts, See Table No minimum for single outputs models See Table for multi outputs models 3 s max 12 ms 10 ms minimum at 110 VAC 1% 0.5% max at full load See Table 4% max deviation, 500 s recovery time for a 25% load change 1% max pk-pk (See Note 1) Provided on output 1 only; set at 112-132% of its nominal output voltage 125-200% with auto recovery Trip & restart with auto recovery 0.04%/C
Environmental
Operating Temperature Storage Temperature Cooling Relative Humidity Operating Altitude Shock Vibration * 0 C to +70 C, derate linearly from 100% load at +50C to 50% load at +70 C -40 C to +85 C Convection cooled 5-95%, non-condensing 3000m 20 G in 3 directions, 3 times each 0.15mm over 10-55Hz and 1 octave/min. 3 axes for 30 min
* * * * * *
EMC & Safety
EMC ESD Radiated Immunity EFT/Burst Surge Safety Approvals * * * * * * EN55011 (CISPR11) Class B & FCC Class B conducted EN61000-4-2 Level 2 Perf Criteria A EN61000-4-3 3 V/m Perf Criteria A EN61000-4-4 Level 3 Perf Criteria A EN61000-4-5 Level 3 Perf Criteria A UL2601-1, CSA C22.2 No 601.1, EN60601-1, IEC 601-1
* * *
* www.xp-medical.com sales@xp-medical.com
*
OUTPUT VOLTAGE & CURRENT RATINGS
Output Power Maximum 28 W 28 W 28 W 28 W 28 W 25 W 25 W 25 W 25 W 25 W Notes Vnom 5.1 V 12.0 V 15.0 V 24.0 V 28.0 V 5.1 V 5.1 V 5.1 V 5.1 V 5.1 V Output 1 Imin Imax 0.0 A 5.5 A 0.0 A 2.3 A 0.0 A 1.9 A 0.0 A 1.2 A 0.0 A 1.0 A 0.4 A 2.5 A 0.4 A 2.5 A 0.4 A 2.5 A 0.4 A 2.5 A 0.4 A 2.5 A Reg.(2) 2% 1% 1% 1% 1% 3% 3% 3% 3% 3% Vnom Output 2 Imin Imax Reg.(2) Vnom Output 3 Imin Imax Reg.(2)
PBM25
Model Number PBM25US05 PBM25US12* PBM25US15* PBM25US24* PBM25US28* PBM25UD01* PBM25UD02* PBM25UT01* PBM25UT02* PBM25UT03*
+12 +24 +12 +15 +24
V V V V V
0.2 0.1 0.2 0.1 0.1
A A A A A
1.5 0.8 1.5 1.0 0.8
A A A A A
5% 5% 5% 5% 5%
-12 V -15 V -12 V
0.05 A 0.05 A 0.05 A
0.2 A 0.2 A 0.2 A
4% 4% 4%
*
1. Ripple and noise measured with 20 MHz bandwidth with 10 F electrolytic in parallel with 0.1 F ceramic. 2. Regulation includes line and load regulation and set point accuracy. All multiple output models may be operated at no-load, but output regulation will increase to 10% on outputs V2 & V3. * Also available from Farnell InOne, see pages 350 & 351.
Mechanical Details - Single Output Models
PIN CONNECTIONS Pin Function 1 Return 2 Return 3 Output 1 4 Output 1 Notes: 1. Dimensions shown in inches (mm). 2. Tolerance 0.03 (0.76) maximum. 3. Input connector P1 mates with Molex housing 09-50-3031 and Molex 2878 series crimp terminal. 4. Output connector P2 mates with Molex housing 09-50-3041 and Molex 2878 series crimp terminal. 5. Weight 163 g. 6. Covers available, order part number PBM25 COVER*, or add suffix `-C' to have cover fitted (78.8 x 12.0 x 34.8 mm). 7. Mating connector kit available, order part number PBM25 CONKT-S*.
Mechanical Details - Multi Output Models
PIN CONNECTIONS Pin Dual Output Triple Output 1 Output 2 Output 2 2 Output 1 Output 1 3 Output 1 Output 1 4 Common Return Common Return 5 Common Return Common Return 6 No Connection Output 3 Notes: 1. Dimensions shown in inches (mm). 2. Tolerance 0.03 (0.76) maximum. 3. Input connector P1 mates with Molex housing 09-50-3031 and Molex 2878 series crimp terminal. 4. Output connector P2 mates with Molex housing 09-50-3061 and Molex 2878 series crimp terminal. 5. Weight 175 g. 6. Covers available, order part number PBM25 COVER*, or add suffix `-C' to have cover fitted (78.8 x 120.0 x 34.8 mm). 7. Mating connector kit available, order part number PBM25 CONKT-D/T*.
www.xp-medical.com
sales@xp-medical.com
Rev 3.1


▲Up To Search▲   

 
Price & Availability of PBM25UD01

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X